基于APD读出的长CsI(Tl)闪烁体晶体的性能恢复
仪器与探测器
2020-03-30 v2 核实验
摘要
CALIFA是FAIR上R3B实验的高效率、高能量分辨率量能器,用于在散射实验中探测高能轻带电粒子和伽马射线,并于2018至2020年间在FAIR的Phase-0实验期间进行调试。它以分段构型环绕反应靶,由2432个由长CsI(Tl)指形闪烁体晶体构成的探测单元组成。CALIFA作为R3B量能器预期运行寿命为10年,有必要采取措施以确保持久性能。本文中,我们对运行超过四年后的两组各6个不同CALIFA探测器探测单元进行了系统研究。在不同条件下评估了能量分辨率和光输出产额。测试涵盖首批组装探测单元的老化,并研究了退化探测单元的恢复程序。给出了所观测到的退化的可能原因,指向晶体-APD耦合。
引用
@article{arxiv.2003.01485,
title = {Performance recovery of long CsI(Tl) scintillator crystals with APD-based readout},
author = {P. Cabanelas and D. González and H. Alvarez-Pol and J. M. Boillos and E. Casarejos and J. Cederkall and D. Cortina and M. Feijoo and D. Galaviz and E. Galiana and R. Gernhäuser and P. Golubev and A. -L. Hartig and P. Klenze and A. Knyazev and T. Kröll and E. Nácher and J. Park and A. Perea and B. Pietras and L. Ponnath and H. -B. Rhee and J. L. Rodríguez-Sánchez and C. Suerder and O. Tengblad and P. Teubig},
journal= {arXiv preprint arXiv:2003.01485},
year = {2020}
}