English

Single-shot X-ray Absorption Spectroscopy at X-ray Free Electron Lasers

Instrumentation and Detectors 2020-05-05 v1 Materials Science

Abstract

X-ray Absorption Spectroscopy (XAS) is a widely used X-ray diagnostic method. While synchrotrons have large communities of XAS users, its use on X-Ray Free Electron Lasers (XFEL) facilities has been rather limited. At a first glance, the relatively narrow bandwidth and the highly fluctuating spectral structure of XFEL sources seem to prevent high-quality XAS measurements without accumulating over many shots. Here, we demonstrate for the first time the collection of single-shot XAS spectra on an XFEL, with error bars of only a few percent, over tens of eV. We show how this technique can be extended over wider spectral ranges towards Extended X-ray Absorption Fine Structure (EXAFS) measurements, by concatenating a few tens of single-shot measurements. Such results open indisputable perspectives for future femtosecond time resolved XAS studies, especially for transient processes that can be initiated at low repetition rate.

Keywords

Cite

@article{arxiv.2005.01572,
  title  = {Single-shot X-ray Absorption Spectroscopy at X-ray Free Electron Lasers},
  author = {M. Harmand and M. Cammarata and M. Chollet and A. G Krygier and H. T. Lemke and D. Zhu},
  journal= {arXiv preprint arXiv:2005.01572},
  year   = {2020}
}

Comments

11 pages, 5 figures