English

Operation of graphene quantum Hall resistance standard in a cryogen-free table-top system

Mesoscale and Nanoscale Physics 2015-08-27 v1

Abstract

We demonstrate quantum Hall resistance measurements with metrological accuracy in a small cryogen-free system operating at a temperature of around 3.8K and magnetic fields below 5T. Operating this system requires little experimental knowledge or laboratory infrastructure, thereby greatly advancing the proliferation of primary quantum standards for precision electrical metrology. This significant advance in technology has come about as a result of the unique properties of epitaxial graphene on SiC.

Keywords

Cite

@article{arxiv.1507.04601,
  title  = {Operation of graphene quantum Hall resistance standard in a cryogen-free table-top system},
  author = {T. J. B. M. Janssen and S. Rozhko and I. Antonov and A. Tzalenchuk and J. M. Williams and Z. Melhem and H. He and S. Lara-Avila and S. Kubatkin and R. Yakimova},
  journal= {arXiv preprint arXiv:1507.04601},
  year   = {2015}
}

Comments

15 pages, 9 figures