The ANATOMIX beamline at Synchrotron SOLEIL, operational since 2018, is dedicated to hard X-ray full-field tomography techniques. Operating in a range of photon energies from approximately 5 to 50 keV, it offers both parallel-beam projection microtomography, in absorption and phase contrast, and nanotomography using a zone-plate transmission X-ray microscope. With these methods, the beamline covers a range of spatial resolution from 20 nm to 20 μm, expressed in terms of useful pixel size. The variable beam size of up to 40 mm allows users to image large objects. Here we describe the microtomography instrumentation of the beamline.
@article{arxiv.2002.03242,
title = {Microtomography on the ANATOMIX beamline at Synchrotron SOLEIL},
author = {T. Weitkamp and M. Scheel and J. Perrin and G. Daniel and A. King and Le Roux and V. and J. -L. Giorgetta and A. Carcy and F. Langlois and K. Desjardins and C. Menneglier and M. Cerato and C. Engblom and G. Cauchon and T. Moreno and C. Rivard and Y. Gohon and F. Polack},
journal= {arXiv preprint arXiv:2002.03242},
year = {2023}
}
Comments
6 pages, 3 figures. Paper submitted for publication in the Proceedings of the 14th International Conference on Synchrotron Radiation Instrumentation (SRI 2021), 28 March-1 April 2022, Hamburg, Germany, J. Phys. Conf. Series