灵活 A 扫速率 MHz OCT:通过相干平均实现计算降速
医学物理
2016-07-21 v1 光学
摘要
为了实现线速率可调的快速 OCT 系统,我们研究了基于 FDML 的 MHz-OCT 系统图像数据的平均方法。线速率可在软件中降低,以换取系统灵敏度和图像质量的提升。我们比较了相干平均与非相干平均,在后期处理中将 3.2 MHz FDML OCT 系统的速度有效降至约 100 kHz。我们证明,在无需特殊干涉仪设计或数字相位稳定的情况下,MHz 系统即可实现相干平均。我们展示了体内人指关节的高质量、深穿透 OCT 图像。
引用
@article{arxiv.1607.06003,
title = {Flexible A-scan rate MHz OCT: Computational downscaling by coherent averaging},
author = {Tom Pfeiffer and Wolfgang Wieser and Thomas Klein and Markus Petermann and Jan-Phillip Kolb and Matthias Eibl and Robert Huber},
journal= {arXiv preprint arXiv:1607.06003},
year = {2016}
}
备注
This work was supported by the German Research Foundation (DFG) (Hu 1006/6-1) and European Union projects FDML-Raman (FP7 ERC StG, contract no. 259158) and ENCOMOLE-2i (Horizon 2020, ERC CoG no. 646669). in Proc. SPIE 9697, Optical Coherence Tomography and Coherence Domain Optical Methods in Biomedicine XX, 969703 (March 8, 2016)