English

Characterising a Si(Li) detector element for the SIXA X-ray spectrometer

Instrumentation and Detectors 2009-10-30 v1

Abstract

The detection efficiency and response function of a Si(Li) detector element for the SIXA spectrometer have been determined in the 500 eV to 5 keV energy range using synchrotron radiation emitted at a bending magnet of the electron storage ring BESSY, which is a primary radiation standard. The agreement between the measured spectrum and the model calculation is better than 2%. PACS: 95.55.Ka; 07.85.Nc; 29.40.Wk; 85.30.De Keywords: Si(Li) detectors, X-ray spectrometers, detector calibration, X-ray response, spectral lineshape

Keywords

Cite

@article{arxiv.physics/9703004,
  title  = {Characterising a Si(Li) detector element for the SIXA X-ray spectrometer},
  author = {T. Tikkanen and S. Kraft and F. Scholze and R. Thornagel and G. Ulm},
  journal= {arXiv preprint arXiv:physics/9703004},
  year   = {2009}
}

Comments

11 pages, 11 PostScript figures, uses elsart.sty, submitted to Nucl. Instrum. Meth. A

R2 v1 2026-07-22T19:16:35.127Z