Electrode thickness measurement of a Si(Li) detector for the SIXA array

Instrumentation and Detectors 2009-10-30 v1

Abstract

Cathode electrodes of the Si(Li) detector elements of the SIXA X-ray spectrometer array are formed by gold-palladium alloy contact layers. The equivalent thickness of gold in one element was measured by observing the characteristic L-shell X-rays of gold excited by monochromatised synchrotron radiation with photon energies above the L3 absorption edge of gold. The results obtained at 4 different photon energies below the L2 edge yield an average value of 22.4(35) nm which is consistent with the earlier result extracted from detection efficiency measurements. PACS: 29.40.Wk; 85.30.De; 07.85.Nc; 95.55.Ka Keywords: Si(Li) detectors, X-ray spectrometers, X-ray fluorescence, detector calibration, gold electrodes, synchrotron radiation

Keywords

Cite

@article{arxiv.physics/9710009,
  title  = {Electrode thickness measurement of a Si(Li) detector for the SIXA array},
  author = {T. Tikkanen and K. Hamalainen and S. Huotari},
  journal= {arXiv preprint arXiv:physics/9710009},
  year   = {2009}
}

Comments

10 pages, 4 PostScript figures, uses elsart.sty, submitted to Nucl. Instrum. Meth. A