Abstract
Accurate magnetic field direction sensing in compact platforms is critical in applications spanning magnetic navigation, space science, and biomedical imaging. We demonstrate a single-optical-axis vector optically pumped magnetometer based on Rabi oscillations between Zeeman sublevels driven by a series of resonant radiofrequency (RF) polarization ellipses (PEs). A calibration protocol based on controlled rotations of the DC magnetic field determines the spatial orientation of each PE. We develop a detailed theoretical model describing the angular dependence of the Rabi frequencies, incorporating key systematics including RF Stark shifts and Bloch-Siegert shifts. We also account for an RF-based heading-error systematic affecting Rabi-frequency measurements arising from the nonlinear Zeeman effect. Simultaneous Larmor measurements yield the magnitude of the magnetic field, enabling integrated vector-scalar measurements. The magnetometer achieves deadzone-free vector operation with 80 rad mean angular accuracy and angular noise densities as low as 8 rad, offering a pathway towards miniaturized sensors without requiring 3D optical access or sensor rotations.
Keywords
Cite
@article{arxiv.2603.08594,
title = {An Accurate Vector Magnetometer via Zeeman Rabi Oscillations},
author = {Thanmay S. Menon and Dawson P. Hewatt and Christopher Kiehl and Michaela Ellmeier and Svenja Knappe and Cindy A. Regal},
journal= {arXiv preprint arXiv:2603.08594},
year = {2026}
}