Abstract
We have designed and tested an atomic vectorial magnetometer based on the analysis of the coherent oscillatory transients in the transmission of resonant laser light through a Rb vapor cell. We show that the oscillation amplitudes at the Larmor frequency and its first harmonic are related through a simple formula to the angles determining the orientation of the magnetic field vector. The magnetometer was successfully applied to the measurement of the ambient magnetic field.
Keywords
Cite
@article{arxiv.1402.5540,
title = {Vectorial atomic magnetometer based on coherent transients of laser absorption in Rb vapor},
author = {L. Lenci and A. Auyuanet and S. Barreiro and P. Valente and A. Lezama and H. Failache},
journal= {arXiv preprint arXiv:1402.5540},
year = {2015}
}
Comments
6 pages, 6 figures