English

Uniform non-stoichiometric titanium nitride thin films for improved kinetic inductance detector array

Instrumentation and Detectors 2016-02-17 v2

Abstract

We describe the fabrication of homogeneous sub-stoichiometric titanium nitride films for microwave kinetic inductance detector (mKID) arrays. Using a 6 inch sputtering target and a homogeneous nitrogen inlet, the variation of the critical temperature over a 2 inch wafer was reduced to <25 %. Measurements of a 132-pixel mKID array from these films reveal a sensitivity of 16 kHz/pW in the 100 GHz band, comparable to the best aluminium mKIDs. We measured a noise equivalent power of NEP = 3.6e-15 W/Hz^(1/2). Finally, we describe possible routes to further improve the performance of these TiN mKID arrays.

Keywords

Cite

@article{arxiv.1510.01876,
  title  = {Uniform non-stoichiometric titanium nitride thin films for improved kinetic inductance detector array},
  author = {G. Coiffard and K-F. Schuster and E. F. C. Driessen and S. Pignard and M. Calvo and A. Catalano and J. Goupy and A. Monfardini},
  journal= {arXiv preprint arXiv:1510.01876},
  year   = {2016}
}

Comments

7 pages, 4 figures, submitted to Journal of low temperature physics, Proceedings of LTD-16

R2 v1 2026-06-22T11:14:38.781Z