Time resolution and characteristic study of MWPC detectors with different Argon based gas

Instrumentation and Detectors 2021-10-06 v1 High Energy Physics - Experiment Nuclear Experiment

Abstract

A prototype of Multi-Wire Proportional Chambers (MWPC) has been fabricated for the study of its various characteristics. The detector contains gold-coated tungsten wires (20 μm\mu m diameter) on the anode frame, with a pitch of 2.8 mm. The gap between the anode and the cathode is 3 mm and the gap between anode and read-out is also 3 mm. Detailed study of MWPC in terms of gain, energy and timing resolution and efficiency measurements have been performed. The detector has been operated using Ar/CO2_{2} gas mixtures with 70:30 and 90:10 ratio. Energy spectrum of 55^{55}Fe X-ray source is obtained for the detector. The gain and energy resolution of the detector were calculated using X-ray spectrum. Time resolution is obtained \sim10 ns.

Keywords

Cite

@article{arxiv.1810.03096,
  title  = {Time resolution and characteristic study of MWPC detectors with different Argon based gas},
  author = {Rajendra Nath Patra and R. N. Singaraju and S. Biswas and T. K. Nayak and Y. P. Viyogi},
  journal= {arXiv preprint arXiv:1810.03096},
  year   = {2021}
}

Comments

3 pages, 7 figures