English

Performance of a Time-Projection-Chamber with a Large-Area Micro-Pixel-Chamber Readout

Instrumentation and Detectors 2008-11-26 v1

Abstract

A micro time-projection-chamber (micro-TPC) with a detection volume of 23*28*31 cm^3 was developed, and its fundamental performance was examined. The micro-TPC consists of a micro pixel chamber with a detection area of 31*31 cm^2 as a two-dimensional imaging device and a gas electron multiplier with an effective area of 23*28 cm^2 as a pre-gas-multiplier. The micro-TPC was operated at a gas gain of 50,000, and energy resolutions and spatial resolutions were measured.

Keywords

Cite

@article{arxiv.physics/0701085,
  title  = {Performance of a Time-Projection-Chamber with a Large-Area Micro-Pixel-Chamber Readout},
  author = {Kentaro Miuchi and Kaori Hattori and Shigeto Kabuki and Hidetoshi Kubo and Shunsuke Kurosawa and Hironobu Nishimura and Atsushi Takada and Ken'ichi Tsuchiya and Yoko Okada and Toru Tanimori and Kazuki Ueno},
  journal= {arXiv preprint arXiv:physics/0701085},
  year   = {2008}
}

Comments

4 pages, 7 figures, proceedings of IWORID 8