English

Multigap RPC time resolution to 511 keV annihilation photons

Instrumentation and Detectors 2015-12-09 v1 Medical Physics

Abstract

The time resolution of Multigap Resistive Plate Counters (MRPCs) to 511511 keV gamma rays has been investigated using a 22^{22}Na source and four detectors. The MRPCs time resolution has been derived from the Time-of-Flight information, measured from pairs of space correlated triggered events. A GEANT4 simulation has been performed to analyze possible setup contributions and to support experimental results. A time resolution (FWHM) of 376376 ps and 312312 ps has been measured for a single MRPC with four 250250 μ\mum gas gaps by considering respectively one and two independent pairs of detectors.

Keywords

Cite

@article{arxiv.1406.2119,
  title  = {Multigap RPC time resolution to 511 keV annihilation photons},
  author = {G. Belli and M. Gabusi and G. Musitelli and R. Nardò and S. P. Ratti and A. Tamborini and P. Vitulo},
  journal= {arXiv preprint arXiv:1406.2119},
  year   = {2015}
}

Comments

25 pages, 14 figures