Spatio-angular fluorescence microscopy II. Paraxial 4$f$ imaging

Optics 2019-07-16 v1 Quantitative Methods

Abstract

We investigate the properties of a single-view fluorescence microscope in a 4ff geometry when imaging fluorescent dipoles without using the monopole or scalar approximations. We show that this imaging system has a spatio-angular band limit, and we exploit the band limit to perform efficient simulations. Notably, we show that information about the out-of-plane orientation of ensembles of in-focus fluorophores is recorded by paraxial fluorescence microscopes. Additionally, we show that the monopole approximation may cause biased estimates of fluorophore concentrations, but these biases are small when the sample contains either many randomly oriented fluorophores in each resolvable volume or unconstrained rotating fluorophores.

Keywords

Cite

@article{arxiv.1901.01181,
  title  = {Spatio-angular fluorescence microscopy II. Paraxial 4$f$ imaging},
  author = {Talon Chandler and Hari Shroff and Rudolf Oldenbourg and Patrick La Rivière},
  journal= {arXiv preprint arXiv:1901.01181},
  year   = {2019}
}

Comments

21 pages, 12 figures. Portions of this work previously appeared as arXiv:1812.07093, which was split during refereeing