English

Readout of GEM Detectors Using the Medipix2 CMOS Pixel Chip

Instrumentation and Detectors 2008-11-26 v1

Abstract

We have operated a Medipix2 CMOS readout chip, with amplifying, shaping and charge discriminating front-end electronics integrated on the pixel-level, as a highly segmented direct charge collecting anode in a three-stage gas electron multiplier (Triple-GEM) to detect the ionization from 55^{55}Fe X-rays and electrons from 106^{106}Ru. The device allows to perform moderate energy spectroscopy measurements (20 % FWHM at 5.9 keV XX-rays) using only digital readout and two discriminator thresholds. Being a truly 2D-detector, it allows to observe individual clusters of minimum ionizing charged particles in Ar/CO2Ar/CO_2 (70:30) and He/CO2He/CO_2 (70:30) mixtures and to achieve excellent spatial resolution for position reconstruction of primary clusters down to 50μm\sim 50 \mu m, based on the binary centroid determination method.

Keywords

Cite

@article{arxiv.physics/0611229,
  title  = {Readout of GEM Detectors Using the Medipix2 CMOS Pixel Chip},
  author = {A. Bamberger and K. Desch and U. Renz and M. Titov and N. Vlasov and P. Wienemann and A. Zwerger},
  journal= {arXiv preprint arXiv:physics/0611229},
  year   = {2008}
}

Comments

18 pages, 14 pictures. submitted to Nuclear Instruments and Methods in Physics Research A

R2 v1 2026-07-22T19:13:49.695Z