English

PMT signal increase using a wavelength shifting paint

Instrumentation and Detectors 2015-12-09 v1 Nuclear Experiment

Abstract

We report a 1.65 times increase of the PMT signal and a simple procedure of application of a new wavelength shifting (WLS) paint for PMTs with non-UV-transparent windows. Samples of four different WLS paints, made from hydrocarbon polymers and organic fluors, were tested on a 5-inch PMT (ET 9390KB) using Cherenkov radiation produced in fused silica disks by 106^{106}Ru electrons on a `table-top' setup. The best performing paint was employed on two different types of 5-inch PMTs (ET 9390KB and XP4572B), installed in atmospheric pressure CO2_2 gas Cherenkov detectors, and tested using GeV electrons.

Keywords

Cite

@article{arxiv.1502.01772,
  title  = {PMT signal increase using a wavelength shifting paint},
  author = {K. Allada and Ch. Hurlbut and L. Ou and B. Schmookler and A. Shahinyan and B. Wojtsekhowski},
  journal= {arXiv preprint arXiv:1502.01772},
  year   = {2015}
}

Comments

6 pages, 8 figures, accepted NIM-A

R2 v1 2026-06-22T08:23:26.969Z