Inter-laboratory nanoamp current comparison with sub-part-per-million uncertainty

Instrumentation and Detectors 2019-03-08 v1

Abstract

An ultrastable low-noise current amplifier (ULCA) has been transferred between two laboratories, NPL and PTB, four times, with three of the transfers yielding a relative change in the trans-resistance gain of less than 2×1072 \times 10^{-7}. This is a new bench-mark for inter-laboratory transfer of small current. We describe in detail the use of a cryogenic current comparator to calibrate the ULCA at NPL, and the use of the ULCA to measure 11 GΩ\Omega resistors with relative uncertainties at the 10710^{-7} level.