Comparison between Thermoelectric and Bolometric Microwave Power Standards

Instrumentation and Detectors 2021-02-24 v1 Classical Physics

Abstract

In the paper, a comparison is described of the microwave power standard based on thermoelectric sensors against an analogous standard based on bolometric sensors. Measurements have been carried out with the classical twin-type microcalorimeter, fitted with N-connector test ports suitable for the frequency band 0.05 - 18 GHz. An appropriate measurand definition is given for being suitable to both standard types. A system accuracy assessment is performed applying the Gaussian error propagation through the mathematical models that interpret the microcalorimeter response in each case. The results highlight advantages and weaknesses of each power standard type.

Keywords

Cite

@article{arxiv.2102.11607,
  title  = {Comparison between Thermoelectric and Bolometric Microwave Power Standards},
  author = {Luciano Brunetti and Luca Oberto and Marco Sellone and Emil Vremera},
  journal= {arXiv preprint arXiv:2102.11607},
  year   = {2021}
}

Comments

6 pages, 6 figures, 3 tables