English

Ultra-low vibration closed-cycle cryogenic surface-electrode ion trap apparatus

Instrumentation and Detectors 2021-06-10 v2 Quantum Physics

Abstract

We describe the design, commissioning and operation of an ultra-low-vibration closed-cycle cryogenic ion trap apparatus. One hundred lines for low-frequency signals and eight microwave / radio frequency coaxial feed lines offer the possibility of implementing a small-scale ion-trap quantum processor or simulator. With all supply cables attached, more than 1.3 W of cooling power at 5 K is still available for absorbing energy from electrical pulses introduced to control ions. The trap itself is isolated from vibrations induced by the cold head using a helium exchange gas interface. The performance of the vibration isolation system has been characterized using a Michelson interferometer, finding residual vibration amplitudes on the order of 10 nm rms. Trapping of 9^9Be+^+ ions has been demonstrated using a combination of laser ablation and photoionization.

Keywords

Cite

@article{arxiv.2008.05601,
  title  = {Ultra-low vibration closed-cycle cryogenic surface-electrode ion trap apparatus},
  author = {Timko Dubielzig and Sebastian Halama and Henning Hahn and Giorgio Zarantonello and Malte Niemann and Amado Bautista-Salvador and Christian Ospelkaus},
  journal= {arXiv preprint arXiv:2008.05601},
  year   = {2021}
}

Comments

10 figures