Towards low-cost lead screening with transmission XRF

Instrumentation and Detectors 2025-11-13 v1

Abstract

Human exposure to lead (Pb) is a global health concern, yet existing technologies for detecting lead in our environment remain prohibitively expensive for widespread deployment. Here we present a new concept towards lead screening using X-ray fluorescence (XRF) in an unconventional geometry we coin transmission XRF in which the sample is placed between the source and detector. For cost reduction, we then show that 241^{241}Am found in ionizing smoke detectors is spectrally suitable for Pb L-shell XRF generation and can thus replace X-ray tubes used in conventional XRF devices. Exploring soil screening as the first application, we demonstrate with Monte Carlo simulations that a configuration with 7×\times 241^{241}Am sources and a standard silicon drift detector can enable screening-relevant detection limits (100 ppm Pb) in soil within practical measurement times (<30 min). We believe this concept opens a route toward low-cost and scalable XRF instrumentation for democratizing lead screening across a wide range of samples.

Keywords

Cite

@article{arxiv.2511.09110,
  title  = {Towards low-cost lead screening with transmission XRF},
  author = {Christoph Gaßner and Juliane Reisewitz and Jenna E. Forsyth and Kian Shaker},
  journal= {arXiv preprint arXiv:2511.09110},
  year   = {2025}
}

Comments

9 pages w/ 4 figures (main) + 4 pages w/ 1 figure (supplementary). Released as a part of OpenXRF (https://openxrf.org/)