English

Spatial resolution determination of a position sensitive ultra-cold neutron detector

Instrumentation and Detectors 2022-07-27 v1 High Energy Physics - Experiment

Abstract

The study of the properties of the quantum states of bouncing neutrons requires position sensitive detection with micro-metric spatial resolution. The UCNBoX detector relies on Charge Coupled Devices (CCD) coated with a thin boron-10 conversion layer to detect neutron hits. In this paper, we present an original experimental method to determine the spatial resolution of this device using micrometric masks. The observed resolution is 2.0±0.3 μ2.0\pm0.3~\mum.

Keywords

Cite

@article{arxiv.2203.09335,
  title  = {Spatial resolution determination of a position sensitive ultra-cold neutron detector},
  author = {B. Clément and L. Gesson and T. Jenke and V. V. Nesvizhevsky and G. Pignol and S. Roccia and J. -P. Scordillis},
  journal= {arXiv preprint arXiv:2203.09335},
  year   = {2022}
}

Comments

4 pages, 4 figure Vienna Conference on Instrumentation 2022

R2 v1 2026-06-24T10:17:08.278Z