Abstract

A multi-reflection time-of-flight mass spectrograph, competitive with Penning trap mass spectrometers, has been built at RIKEN. We have performed a first online mass measurement, using 8Li+ (T1/2 = 838 ms). A new analysis method has been realized, with which, using only 12C+ references, the mass excess of 8Li was accurately determined to be 20947.6(15)(34) keV (dm/m = 6.6 x 10-7). The speed, precision and accuracy of this first online measurement exemplifies the potential for using this new type of mass spectrograph for precision measurements of short-lived nuclei.

Cite

@article{arxiv.1307.7373,
  title  = {Single-reference high-precision mass measurement with a multi-reflection time-of-flight mass spectrograph},
  author = {Y. Ito and P. Schury and M. Wada and S. Naimi and T. Sonoda and H. Mita and F. Arai and A. Takamine and K. Okada and A. Ozawa and H. Wollnik},
  journal= {arXiv preprint arXiv:1307.7373},
  year   = {2015}
}

Comments

6 pages, 4 figures