English

Simulating the secondary electron avalanche of MCP by Geant4

Instrumentation and Detectors 2024-05-30 v3

Abstract

Nowadays, Microchannel Plate (MCP), as a kind of electron multipliers based on the secondary electron emission, is widely used in many high-sensitive experiments, such as neutrino detection, which require the noise to be as low as possible, while the conventional straight-channel MCP will inevitably have ion feedback, resulting in the sequential after-pulses being the major source of noise. Normally, the problem can be effectively avoided by coupling two straight-channel MCPs in cascade and combining the channels into a `V` shape known as chevron MCPs, but this method is limited by the manufacturing techniques due to the unavoidable gap between the two pieces that will worsen the resolution and peak-to-valley ratio. However, the ion feedback can be inhibited significantly for MCPs with curved channels. Based on Geant4, we investigate how the geometrical parameters of curved-channel MCP influence the gain and get the optimum pore diameter for an MCP to reach the maximum gain with fixed thickness and applied voltage. Additionally, the track-by-track simulation reveals that the average acceleration distance of a secondary electron inside the curved-channel is approximately 20 um when the applied voltage, length-to-diameter ratio and pore diameter are 950 V, 50:1 and 20 um, respectively.

Cite

@article{arxiv.2310.07153,
  title  = {Simulating the secondary electron avalanche of MCP by Geant4},
  author = {Huaxing Peng and Baojun Yan and Han Miao and Shulin Liu and Binting Zhang},
  journal= {arXiv preprint arXiv:2310.07153},
  year   = {2024}
}

Comments

29 pages, 14 figures

R2 v1 2026-06-28T12:46:50.784Z