English

Shot-noise limited, supercontinuum based optical coherence tomography

Optics 2022-10-18 v1

Abstract

We present the first demonstration of shot-noise limited supercontinuum-based spectral domain optical coherence tomography (SD-OCT) with axial resolution of 5.9 μ\mum at a center wavelength of 1370 nm. Current supercontinuum-based SD-OCT systems cannot be operated in the shot-noise limited detection regime because of severe pulse-to-pulse relative intensity noise of the supercontinuum source. To overcome this disadvantage we have developed a low-noise supercontinuum source based on an all-normal dispersion (ANDi) fiber, pumped by a femtosecond laser. The noise performance of our 90 MHz ANDi supercontinuum source is compared to that of two commercial sources operating at 80 and 320 MHz repetition rate. We show that the low noise of the ANDi supercontinuum source improves the OCT images significantly in terms of both higher contrast, better sensitivity, and improved penetration. From SD-OCT imaging of skin, retina, and multi-layer stacks we conclude that supercontinuum-based SD-OCT can enter the domain of shot-noise limited detection.

Keywords

Cite

@article{arxiv.2010.05226,
  title  = {Shot-noise limited, supercontinuum based optical coherence tomography},
  author = {Shreesha Rao D. S. and Mikkel Jensen and Lars Grüner-Nielsen and Jesper Toft Olsen and Peter Heiduschka and Björn Kemper and Jürgen Schnekenburger and Martin Glud and Mette Mogensen and Niels Møller Israelsen and Ole Bang},
  journal= {arXiv preprint arXiv:2010.05226},
  year   = {2022}
}

Comments

14 pages, 6 combined images

R2 v1 2026-06-23T19:14:59.913Z