English

Recovery of HADES drift chambers suffering from Malter-like effects

Instrumentation and Detectors 2024-03-20 v1

Abstract

The central tracking system of the HADES detector, installed at the SIS-18 synchrotron at GSI/Darmstadt (Germany), employs large-area, low-mass drift chambers, featuring Aluminum potential wires and small cell sizes. The chambers in front of the magnetic field, closest to the interaction point, have developed significant self-sustained currents and discharges during operation, most probably triggered by isobutane-based gas mixtures. Only both, (i) replacing isobutane by CO2 and (ii) adding 1000 to 3500 ppmv of water into the Ar/CO2 counting gas mixture, individually optimized for a given chamber, allowed to recover the chambers, enabling stable operation in several production runs since then, e.g. with high-intensity heavy-ion induced reactions. The origin of the instability was found to be deposits on the cathode wires, provoking the Malter-like effects, by visual inspection and energy-dispersive X-ray spectroscopy. The charge on the wires accumulated during their lifetime does not point to so-called classical aging, but presumably the interaction of isobutane with materials in the gas flow, residual impurities, and reaction products formed in plasma, e.g., built by discharges.

Keywords

Cite

@article{arxiv.2403.12178,
  title  = {Recovery of HADES drift chambers suffering from Malter-like effects},
  author = {Christian Wendisch and Christian Muentz and Luis Lopes and Erwin Schwab and Joachim Stroth},
  journal= {arXiv preprint arXiv:2403.12178},
  year   = {2024}
}

Comments

3rd International Conference on Detector Stability and Aging Phenomena in Gaseous Detectors 2024 at CERN