Quantitative powder diffraction using a (2+3) surface diffractometer and an area detector

Applied Physics 2022-01-27 v2

Abstract

X-ray diffractometers primarily designed for surface x-ray diffraction are often used to measure the diffraction from powders, textured materials, and fiber-texture samples in so-called 2θ2\theta scans. Unlike high-energy powder diffraction only a fraction of the powder rings is typically measured and the data consists of many detector images across the 2θ2\theta range. Such diffractometers typically scan in directions not possible on a conventional lab-diffractometer, which gives enhanced control of the scattering vector relative to the sample orientation. There are, however, very few examples where the measured intensity is directly used, such as for profile/Rietveld refinement, as is common with other powder diffraction data. Although the underlying physics is known, converting the data is time-consuming and the appropriate corrections are dispersed across several publications, often not with powder diffraction in mind. In this paper we present the angle calculations and correction factors required to calculate meaningful intensities for 2θ2\theta scans with a (2+3)-type diffractometer and an area detector. We also discuss some of the limitations with respect to texture, refraction, and instrumental resolution, and what kind of information one can hope to obtain.

Keywords

Cite

@article{arxiv.2102.08265,
  title  = {Quantitative powder diffraction using a (2+3) surface diffractometer and an area detector},
  author = {Giuseppe Abbondanza and Alfred Larsson and Francesco Carlá and Edvin Lundgren and Gary S. Harlow},
  journal= {arXiv preprint arXiv:2102.08265},
  year   = {2022}
}

Comments

Research paper, 15 pages, 12 figures, 3 tables

R2 v1 2026-06-23T23:13:04.021Z