Pressurized Helium-4 (PHe) based fast neutron scintillation detector offers an useful alternative to organic liquid-based scintillator due to its relatively low response to the γ-rays compared to the latter type of scintillator. In the present work, we have investigated the capabilities of a PHe detector for the detection of fast neutrons in a mixed radiation field where both the neutrons and the γ-rays are present. Discrimination between neutrons and γ-rays is achieved by using fast-slow charge integration method. We have also conducted systematic studies of the attenuation of fast neutrons and γ-rays by high-density polyethylene (HDPE). Additionally, the simulation analyses, conducted using GEANT4, provide detailed insights into the interactions of the radiation quanta with the PHe detector.