Mini-review on the temporal resolution of fluorescence imaging systems

Medical Physics 2014-07-25 v2 Biological Physics Optics

Abstract

Imaging of rapidly occurring events in biology requires high temporal resolution. In this review article, I have tried to investigate an approximate estimate to quantify temporal resolution limit of a fluorescence imaging system. It was realised that, the temporal resolution is essentially determined by the fact that, once excited, almost all (99.9%99.9\%) excited molecules relax to ground state. We determined the time required for 99.9%99.9\% of molecules to relax is about 3τp=3/(kf+knr)log10e3\tau_p = 3/{(k_f +k_{nr})\log_{10}e}, where kf+knrk_f +k_{nr} is the total emission rate. To arrive at this expression we assumed that, the quantum efficiency of the detector is unity, no scattering and there are no other loses. We further discuss few microscopy technique that are capable of high temporal resolution. These technique includes, multifocal multiphoton microscopy (MMM), multiple excitation spot optical microscopy (MESO) and multiple light-sheet microscopy (MLSM).

Keywords

Cite

@article{arxiv.1407.6189,
  title  = {Mini-review on the temporal resolution of fluorescence imaging systems},
  author = {Partha Pratim Mondal},
  journal= {arXiv preprint arXiv:1407.6189},
  year   = {2014}
}

Comments

8 pages