English

Measurement Of Energy Resolution For TES Microcalorimeter With Optical Pulses

Instrumentation and Detectors 2016-12-08 v1 Nuclear Experiment

Abstract

Energy resolution is an important figure of merit for TES microcalorimeter. We propose a laser system to measure the energy resolution of TES microcalorimeter with a 1550 nm laser source. Compared to method that characterizes the performance by irradiating the detector using X-ray photons from a radioactive source placed inside the refrigerator, our system is safer and more convenient. The feasibility of this system has been demonstrated in the measurement of an Al/Ti bilayer TES microcalorimeter. In this experiment, the tested detector showed a energy resolution of 72 eV in the energy range from 0.2 keV to 0.9 keV

Keywords

Cite

@article{arxiv.1612.02028,
  title  = {Measurement Of Energy Resolution For TES Microcalorimeter With Optical Pulses},
  author = {Shuo Zhang and Qing-Ya Zhang and CongZhan Liu and Jianshe Liu and Wenhui Dong and Wei Chen},
  journal= {arXiv preprint arXiv:1612.02028},
  year   = {2016}
}

Comments

3 page