English

Low-Order Nonlinear Animal Model of Glucose Dynamics for a Bihormonal Intraperitoneal Artificial Pancreas

Optimization and Control 2022-02-02 v2 Other Quantitative Biology

Abstract

Objective: The design of an Artificial Pancreas (AP) to regulate blood glucose levels requires reliable control methods. Model Predictive Control has emerged as a promising approach for glycemia control. However, model--based control methods require computationally simple and identifiable mathematical models that represent glucose dynamics accurately, which is challenging due to the complexity of glucose homeostasis. Methods: In this work, a simple model is deduced to estimate blood glucose concentration in subjects with Type 1 Diabetes Mellitus (T1DM). Novel features in the model are power--law kinetics for intraperitoneal insulin absorption and a separate glucagon sensitivity state. Profile likelihood and a method based on singular value decomposition of the sensitivity matrix are carried out to assess parameter identifiability and guide a model reduction for improving the identification of parameters. Results: A reduced model with 10 parameters is obtained and calibrated, showing good fit to experimental data from pigs where insulin and glucagon boluses were delivered in the intraperitoneal cavity. Conclusion: A simple model with power--law kinetics can accurately represent glucose dynamics submitted to intraperitoneal insulin and glucagon injections. The reduced model was found to exhibit local practical as well as structural identifiability. Importance: The proposed model facilitates intraperitoneal bi-hormonal model-based closed-loop control in animal trials.

Keywords

Cite

@article{arxiv.2011.04636,
  title  = {Low-Order Nonlinear Animal Model of Glucose Dynamics for a Bihormonal Intraperitoneal Artificial Pancreas},
  author = {Claudia Lopez-Zazueta and Øyvind Stavdahl and Anders Lyngvi Fougner},
  journal= {arXiv preprint arXiv:2011.04636},
  year   = {2022}
}