English

A contraction theory approach to observer-based controller design for glucose regulation in type 1 diabetes with intra-patient variability

Systems and Control 2022-06-30 v1 Systems and Control

Abstract

While the Artificial Pancreas is effective in regulating the blood glucose in the safe range of 70-180 mg/dl in type 1 diabetic patients, the high intra-patient variability, as well as exogenous meal disturbances, poses a serious challenge. The existing control algorithms thus require additional safety algorithms and feed-forward actions. Moreover, the unavailability of insulin sensors in Artificial Pancreas makes this task more difficult. In the present work, a subcutaneous model of type 1 diabetes (T1D) is considered for observer-based controller design in the framework of contraction analysis. A variety of realistic multiple-meal scenarios for three virtual T1D patients have been investigated with +30 % and -30 % of parametric variability. The average time spent by the three T1D patients is found to be 77 %, 73 % and 76 %, respectively. A significant reduction in the time spent in hyperglycemia (>180 mg/dl) is achieved without any feed-forward action for meal compensation.

Keywords

Cite

@article{arxiv.2206.14436,
  title  = {A contraction theory approach to observer-based controller design for glucose regulation in type 1 diabetes with intra-patient variability},
  author = {Bhabani Shankar Dey and Anirudh Nath and Abhilash Patel and Indra Narayan Kar},
  journal= {arXiv preprint arXiv:2206.14436},
  year   = {2022}
}