In this work, we have developed and characterized a position-sensitive CLYC detector that acts as the neutron imaging layer and γ-ray Compton scatterer of the novel dual \g-ray and neutron imaging system GN-Vision, which aims at simultaneously obtaining information about the spatial origin of \g-ray and neutron sources. We first investigated the performance of large 50×50~mm2 monolithic CLYC crystals coupled to a pixelated SiPM in terms of energy resolution and neutron-gamma discrimination. The response of two different 95\% 6Li-enriched CLYC detectors coupled to an array of 8×8 SiPMs was studied in comparison to the results of a conventional photo-multiplier tube. Energy resolution ranging from 6-8\% for the 137Cs peak and a figure of merit of 3-4 for the neutron-gamma discrimination have been obtained. The spatial response of the CLYC-SiPM detector to γ-rays and neutrons has also been characterized using charge modulation-based multiplexing techniques based on a diode-coupled charge division circuit. Average resolutions close to 5~mm FWHM with good linearity are obtained in the transverse crystal plane. Last, this work presents the first proof-of-concept experiments of the neutron imaging capability using a neutron pinhole collimator attached to the developed position sensitive CLYC detector.
@article{arxiv.2410.12533,
title = {Imaging neutrons with a position-sensitive monolithic CLYC detector},
author = {J. Lerendegui-Marco and G. Cisterna and J. Hallam and V. Babiano-Suárez and J. Balibrea-Correa and D. Calvo and I. Ladarescu and G. de la Fuente and B. Gameiro and A. Sanchis-Moltó and P. Torres-Sánchez and C. Domingo-Pardo},
journal= {arXiv preprint arXiv:2410.12533},
year = {2025}
}