English

High speed optical holography of retinal blood flow

Instrumentation and Detectors 2016-07-27 v1 Medical Physics Optics

Abstract

We performed non-invasive video imaging of retinal blood flow in a pigmented rat by holographic interferometry of near-infrared laser light backscattered by retinal tissue, beating against an off-axis reference beam sampled at a frame rate of 39 kHz with a high throughput camera. Local Doppler contrasts emerged from the envelopes of short-time Fourier transforms and the phase of autocorrelation functions of holograms rendered by Fresnel transformation. This approach permitted imaging of blood flow in large retinal vessels (30 microns diameter) over 400 by 400 pixels with a spatial resolution of 8 microns and a temporal resolution of 6.5 ms.

Keywords

Cite

@article{arxiv.1607.07800,
  title  = {High speed optical holography of retinal blood flow},
  author = {Mathilde Pellizzari and Manuel Simonutti and Julie Degardin and Jose-Alain Sahel and Mathias Fink and Michel Paques and Michael Atlan},
  journal= {arXiv preprint arXiv:1607.07800},
  year   = {2016}
}

Comments

4 pages, 5 figures, 4 links