High Resolution and High Precision Beam Profile Monitor for Particle Accelerator using linear contact image sensor

Medical Physics 2019-12-23 v2 Accelerator Physics

Abstract

A compact beam-profile monitor was constructed using a linear contact image sensor attached to a plastic scintillator and tested using a 230 MeV proton beam. The results indicate that the beam distribution can be obtained in real-time, and the beam position with a precision of up to 0.03 mm. The compactness and high precision of the device hold considerable potential for it to be used as a beam-profile monitor and offline, daily quality assurance monitor in hadron therapy.

Keywords

Cite

@article{arxiv.1906.10380,
  title  = {High Resolution and High Precision Beam Profile Monitor for Particle Accelerator using linear contact image sensor},
  author = {Tung-Yuan Hsiao and Tzung-Yuang Chen and Huan Niu and Chien-Hsu Chen},
  journal= {arXiv preprint arXiv:1906.10380},
  year   = {2019}
}

Comments

9 pages, 6 figures, 1 table