English

GIOVE - A New Detector Setup for High Sensitivity Germanium Spectroscopy At Shallow Depth

Instrumentation and Methods for Astrophysics 2015-07-14 v1 Nuclear Experiment Instrumentation and Detectors

Abstract

We report on the development and construction of the high-purity germanium spectrometer setup GIOVE (Germanium Inner Outer Veto), recently built and now operated at the shallow underground laboratory of the Max-Planck-Institut f\"ur Kernphysik, Heidelberg. Particular attention was paid to the design of a novel passive and active shield, aiming at efficient rejection of environmental and muon induced radiation backgrounds. The achieved sensitivity level of <100 {\mu}Bq/kg for primordial radionuclides from U and Th in typical {\gamma} ray sample screening measurements is unique among instruments located at comparably shallow depths and can compete with instruments at far deeper underground sites.

Cite

@article{arxiv.1507.03319,
  title  = {GIOVE - A New Detector Setup for High Sensitivity Germanium Spectroscopy At Shallow Depth},
  author = {Gerd Heusser and Marc Weber and Janina Hakenmüller and Matthias Laubenstein and Manfred Lindner and Werner Maneschg and Hardy Simgen and Dominik Stolzenburg and Herbert Strecker},
  journal= {arXiv preprint arXiv:1507.03319},
  year   = {2015}
}

Comments

17 pages, 11 figures

R2 v1 2026-06-22T10:10:28.538Z