English

A Low-noise Germanium Ionization Spectrometer for Low-background Science

Instrumentation and Detectors 2017-01-19 v1

Abstract

Recent progress on the development of very low noise high purity germanium ionization spectrometers has produced an instrument of 1.2 kg mass and excellent noise performance. The detector was installed in a low-background cryostat intended for use in a direct detection search for low mass, WIMP dark matter. This Transaction reports the thermal characterization of the cryostat, specifications of the newly prepared 1.2 kg p-type point contact germanium detector, and the spectroscopic performance of the integrated system. The integrated detector and low background cryostat achieved full-width-at-half-maximum noise performance of 98 eV for an electronic pulse generator peak and 1.9 keV for the 1332 keV Co-60 gamma ray.

Keywords

Cite

@article{arxiv.1603.01584,
  title  = {A Low-noise Germanium Ionization Spectrometer for Low-background Science},
  author = {Craig E. Aalseth and Juan I. Collar and Jim Colaresi and James E. Fast and Todd W. Hossbach and John L. Orrell and Cory T. Overman and Bjorn Scholz and Brent A. VanDevender and K. Michael Yocum},
  journal= {arXiv preprint arXiv:1603.01584},
  year   = {2017}
}

Comments

Submitted to IEEE Transaction in Nuclear Science

R2 v1 2026-06-22T13:04:08.496Z