Magnetic particle imaging (MPI) is a tracer-based technique that directly detects the distribution of magnetic iron-oxide nanoparticles with millisecond temporal resolution and no tissue background. Despite extensive preclinical work, in-vivo application of MPI in humans has not previously been reported. Here, we report the first in-vivo human MPI angiography, visualizing venous perfusion of the upper extremity using a human-scale scanner and clinically approved ferucarbotran. Under identical procedural conditions, we performed X-ray digital subtraction angiography as the clinical gold standard. MPI visualized major superficial and deep veins, including inflow, branching, valve filling, and clearance dynamics in real time with 2 frames per second. These results establish magnetic particle imaging as a clinically translatable modality for radiation-free vascular imaging in humans and mark the transition of MPI from preclinical research to first clinical application.
Cite
@article{arxiv.2603.12010,
title = {First in-vivo human magnetic particle imaging},
author = {Patrick Vogel and Thomas Kampf and Martin A. Rückert and Johanna Günther and Teresa Reichl and Thorsten A. Bley and Volker C. Behr and Philipp Gruschwitz and Viktor Hartung},
journal= {arXiv preprint arXiv:2603.12010},
year = {2026}
}