English

Development of a 5 MHz frequency difference pre-multiplier for a short term frequency stability bench of the oscillators

Instrumentation and Detectors 2007-06-14 v1

Abstract

This paper reports the realization of a 5 MHz frequency difference x10 pre-multiplier, developed in the laboratory to replace an obsolete one. The principle we chose is to synthesize a 45 MHz and a 50 MHz from a reference signal and from the signal to be measured, and to subtract one to the other to generate a 5 MHz, whose precision on the measurement is increased by a factor 10. Obtained Allan variance y() at 1 s is 5.10-14 and output spectral density of phase noise floor is about -160 dBc/Hz at 5 MHz.

Keywords

Cite

@article{arxiv.0706.1853,
  title  = {Development of a 5 MHz frequency difference pre-multiplier for a short term frequency stability bench of the oscillators},
  author = {Patrice Salzenstein and Xavier Jouvenceau and Xavier Vacheret and Gilles Martin and Franck Lardet-Vieudrin},
  journal= {arXiv preprint arXiv:0706.1853},
  year   = {2007}
}