English

Cryogenic environment and performance for testing the Planck radiometers

Instrumentation and Methods for Astrophysics 2010-01-27 v1 Cosmology and Nongalactic Astrophysics

Abstract

This paper is part of the Prelaunch status LFI papers published on JINST: http://www.iop.org/EJ/journal/-page=extra.proc5/jinst The Planck LFI Radiometer Chain Assemblies (RCAs) have been calibrated in two dedicated cryogenic facilities. In this paper the facilities and the related instrumentation are described. The main satellite thermal interfaces for the single chains have to be reproduced and stability requirements have to be satisfied. Setup design, problems occurred and improving solutions implemented are discussed. Performance of the cryogenic setup are reported.

Keywords

Cite

@article{arxiv.1001.4644,
  title  = {Cryogenic environment and performance for testing the Planck radiometers},
  author = {L. Terenzi and M. Lapolla and M. Laaninen and P. Battaglia and F. Cavaliere and A. De Rosa and N. Hughes and P. Jukkala and V. -H. Kilpiä and G. Morgante and M. Tomasi and J. Varis and M. Bersanelli and R. C. Butler and F. Ferrari and C. Franceschet and P. Leutenegger and N. Mandolesi and A. Mennella and R. Silvestri and L. Stringhetti and J. Tuovinen and L. Valenziano and F. Villa},
  journal= {arXiv preprint arXiv:1001.4644},
  year   = {2010}
}

Comments

This is an author-created, un-copyedited version of an article accepted for publication in JINST. IOP Publishing Ltd is not responsible for any errors or omissions in this version of the manuscript or any version derived from it. The definitive publisher authenticated version is available online at 10.1088/1748-0221/4/12/T12015

R2 v1 2026-06-21T14:39:31.101Z