Characteristic study of a quadruple GEM detector and its comparison with a triple GEM detector

Instrumentation and Detectors 2018-10-09 v1 High Energy Physics - Experiment Nuclear Experiment

Abstract

A quadruple GEM detector has been assembled in a standalone configuration and operated using Ar and CO2_2 gas mixtures in proportions of 70:30 and 90:10. Detailed performance study of the detector has been made by using 106^{106}Ru-Rh β\beta-source and X-ray spectrum of 55^{55}Fe source. Results of these measurements are presented in terms of gain, efficiency, energy and time resolutions and also compared with our earlier triple GEM results. The energy resolution has been found to be somewhat worse compared with that of the triple-GEM detector. Effect of drift field on electron transparency and time resolution has been studied in detail.

Keywords

Cite

@article{arxiv.1810.03316,
  title  = {Characteristic study of a quadruple GEM detector and its comparison with a triple GEM detector},
  author = {Rajendra Nath Patra and Rama Narayan Singaraju and Saikat Biswas and Yogendra P. Viyogi and Tapan K. Nayak},
  journal= {arXiv preprint arXiv:1810.03316},
  year   = {2018}
}

Comments

8 pages, 13 figures