In low activity gamma spectrometric measurements of large, dense samples, the bulk sample material shields the HPGe crystal from external background sources. If not accounted for in studies that utilise background-subtraction methods, this effect may result in systematic errors in the sample activity and detection limit estimation. We introduce a Monte Carlo based method to minimise the impact of this effect on sample gamma spectra. It is validated using simulated detector backgrounds and applied to a measurement of low-activity Gd2(SO4)3⋅8H2O. One main prerequisite for the correct application of this method is to know in advance the nuclides which contribute to the detector background spectrum and their spatial distribution. With a thorough understanding of the detector backgrounds, the method improves the accuracy of sensitive low-background measurements of low-activity samples. Even without knowing the background sources and their distribution, conservative results may still be presented that account for the potential systematic errors introduced by this background shielding effect.