An apparatus for studying electrical breakdown in liquid helium at 0.4 K and testing electrode materials for the SNS nEDM experiment

Instrumentation and Detectors 2016-05-10 v2 Nuclear Experiment

Abstract

We have constructed an apparatus to study DC electrical breakdown in liquid helium at temperatures as low as 0.4 K and at pressures between the saturated vapor pressure and \sim600 torr. The apparatus can house a set of electrodes that are 12 cm in diameter with a gap of 121-2 cm between them, and a potential up to ±50\pm 50 kV can be applied to each electrode. Initial results demonstrated that it is possible to apply fields exceeding 100 kV/cm in a 1 cm gap between two electropolished stainless steel electrodes 12 cm in diameter for a wide range of pressures at 0.4 K. We also measured the current between two electrodes. Our initial results, I<1I<1 pA at 45 kV, correspond to a lower bound on the effective volume resistivity of LHe of ρV>5×1018\rho_V > 5\times10^{18} Ω\Omega\cdotcm. This lower bound is 5 times larger than the bound previously measured. We report the design, construction, and operational experience of the apparatus, as well as initial results.

Keywords

Cite

@article{arxiv.1510.06068,
  title  = {An apparatus for studying electrical breakdown in liquid helium at 0.4 K and testing electrode materials for the SNS nEDM experiment},
  author = {T. M. Ito and J. C. Ramsey and W. Yao and D. H. Beck and V. Cianciolo and S. M. Clayton and C. Crawford and S. A. Currie and B. W. Filippone and W. C. Griffith and M. Makela and R. Schmid and G. M. Seidel and Z. Tang and D. Wagner and W. Wei and S. E. Williamson},
  journal= {arXiv preprint arXiv:1510.06068},
  year   = {2016}
}

Comments

15 pages, 14 figures, accepted for publication in Review of Scientific Instruments