English

Absolute measurement of the Fano factor using a Skipper-CCD

Instrumentation and Detectors 2021-07-07 v3

Abstract

Skipper-CCD can achieve deep sub-electron readout noise making possible the absolute determination of the exact number of ionized electrons in a large range, from 0 to above 1900 electrons. In this work we present a novel technique that exploits this unique capability to allow self-calibration and the ultimate determination of silicon properties. We performed an absolute measurement of the variance and the mean number of the charge distribution produced by 55^{55}Fe X-rays, getting a Fano factor absolute measurement in Si at 123K and 5.9 keV. A value of 0.119 ±\pm 0.002 was found and the electron-hole pair creation energy was determined to be (3.749 ±\pm 0.001) eV. This technology opens the opportunity for direct measurements of the Fano factor at low energies.

Keywords

Cite

@article{arxiv.2004.11499,
  title  = {Absolute measurement of the Fano factor using a Skipper-CCD},
  author = {Dario Rodrigues and Kevin Andersson and Mariano Cababie and Andre Donadon and Ana Botti and Gustavo Cancelo and Juan Estrada and Guillermo Fernandez-Moroni and Ricardo Piegaia and Matias Senger and Miguel Sofo Haro and Leandro Stefanazzi and Javier Tiffenberg and Sho Uemura},
  journal= {arXiv preprint arXiv:2004.11499},
  year   = {2021}
}

Comments

six pages, five figures, three tables, one equation