English

The NA62 RICH detector

Instrumentation and Detectors 2017-06-27 v1 High Energy Physics - Experiment

Abstract

The RICH detector of the NA62 experiment at CERN SPS is required to suppress μ+\mu^+ contamination in K+π+ννˉK^+ \to \pi^+ \nu \bar\nu candidate events by a factor at least 100 between 15 and 35 GeV/c momentum, to measure the pion arrival time with 100\sim 100 ps resolution and to produce a trigger for a charged track. It consists of a 17 m long tank filled with Neon gas at atmospheric pressure. \v{C}erenkov light is reflected by a mosaic of 20 spherical mirrors placed at the downstream end of the vessel and is collected by 1952 photomultipliers placed at the upstream end. The construction of the detector will be described and the performance reached during first runs will be discussed.

Keywords

Cite

@article{arxiv.1706.08496,
  title  = {The NA62 RICH detector},
  author = {Andrea Bizzeti},
  journal= {arXiv preprint arXiv:1706.08496},
  year   = {2017}
}

Comments

4 pages, 3 figures

R2 v1 2026-06-22T20:29:58.887Z