Sensitive neutron detection method using delayed coincidence transitions in existing iodine-containing detectors

Instrumentation and Detectors 2017-03-08 v3

Abstract

This letter explains a new, highly sensitive method for the detection of neutrons, which uses the T1/2=845_{1/2}=845 ns delay in the decay of 128^{128}I at the 137.8 keV energy level, resulting from the capture of thermal neutrons by iodine nuclei in NaI and CsI scintillation detectors. The use of delayed coincidence techniques with a several μs\mu {\rm s} time frame for delayed events allows for the highly effective discrimination of neutron events from any existing background signals. A comparison of ambient neutron measurements between those identified through the suggested method from a cylindrical, \o63mm×63mm\, 63 \, {\rm mm}\times 63\, {\rm mm} NaI(Tl) scintillator and those from a low-background proportional 3^3He counter experimentally demonstrates the efficacy of this neutron detection method. For an isotropic, 4π4\pi, thermal neutron flux of 1 ncm2s1{\rm n}\, {\rm cm}^{-2}\, {\rm s}^{-1}, the absolute sensitivity of the NaI detector was found to be 6.5±1countss16.5 \pm 1\, {\rm counts}\, {\rm s}^{-1} with a background of 0.8eventsday10.8\, {\rm events}\, {\rm day}^{-1} and a delay time frame of Δt=1μ\Delta {\rm t}=1\, \mu s. The proposed method can provide low-background experiments, using NaI or CsI, with measurements of the rate and stability of incoming neutron flux to a greater accuracy than 108ncm2s1^{-8}\, {\rm n}\, {\rm cm}^{-2}\, {\rm s}^{-1}.

Keywords

Cite

@article{arxiv.1608.02409,
  title  = {Sensitive neutron detection method using delayed coincidence transitions in existing iodine-containing detectors},
  author = {E. Yakushev and S. Rozov and A. Drokhlyansky and D. Filosofov and Z. Kalaninova and V. Timkin and D. Ponomarev},
  journal= {arXiv preprint arXiv:1608.02409},
  year   = {2017}
}

Comments

accepted for publication / NIM A

R2 v1 2026-06-22T15:14:48.679Z