Abstract

Here we propose double-coil setup to allow high signal-to-noise ratio broad-range heteronuclear magnetic resonance imaging experiments: two independent coils, one of them tuned to 1^{1}H frequency to perform anatomical 1^{1}H imaging, and another one, metamaterial-inspired coil, tuned to the X-nucleus frequency. In this work our goal was to design a broad-range X-nuclei coil to cover 2^{2}H, 11^{11}B, 13^{13}C, 23^{23}Na, 7^{7}Li and 31^{31}P nuclear magnetic resonance frequencies, and to combine it with 1^{1}H coil in one setup. The system was designed for 11.7 T scanner, i.e., with 76-203 MHz frequency tuning range for the X-nuclei and tuned to 500 MHz for the proton coil. X-nuclei coil operates via excitation of the fundamental eigenmode of an array of parallel non-magnetic wires. The excitation of the array is provided via non-resonant feeding loop inductively coupled to the resonator. In order to tune the X-coil over such a wide range, both structural capacitance and inductance of the coil were made variable; narrow range tuning of the 1^{1}H coil is achieved via conventional tuning-matching circuit. Here, the design principle and setup tunability were investigated in simulations and experimentally.