English

Pulse shape simulation for segmented true-coaxial HPGe detectors

Instrumentation and Detectors 2015-03-14 v1 High Energy Physics - Experiment Nuclear Experiment

Abstract

A new package to simulate the formation of electrical pulses in segmented true-coaxial high purity germanium detectors is presented. The computation of the electric field and weighting potentials inside the detector as well as of the trajectories of the charge carriers is described. In addition, the treatment of bandwidth limitations and noise are discussed. Comparison of simulated to measured pulses, obtained from an 18-fold segmented detector operated inside a cryogenic test facility, are presented.

Keywords

Cite

@article{arxiv.1004.0862,
  title  = {Pulse shape simulation for segmented true-coaxial HPGe detectors},
  author = {I. Abt and A. Caldwell and D. Lenz and J. Liu and X. Liu and B. Majorovits},
  journal= {arXiv preprint arXiv:1004.0862},
  year   = {2015}
}

Comments

20 pages, 16 figures

R2 v1 2026-06-21T15:07:01.983Z