English

High voltage scheme optimization for secondary discharge mitigation in GEM-based detectors

Instrumentation and Detectors 2019-10-09 v2

Abstract

We investigate the influence of the high voltage scheme elements on the stability of a detector based on a single 10×1010\times10 cm2^2 area GEM with respect to the secondary discharge occurrence. These violent events pose a major threat to the integrity of GEM detectors and their Front-End Electronics and need to be avoided by any means. For a single GEM setup, we propose a detailed high voltage scheme that is designed to prevent secondary discharges. We determine optimal values of the protection resistors and parasitic capacitances introduced by cables used in the system. The results of this paper may be used as a guideline for the optimization of more complicated multi-GEM detectors.

Keywords

Cite

@article{arxiv.1906.06721,
  title  = {High voltage scheme optimization for secondary discharge mitigation in GEM-based detectors},
  author = {L. Lautner and L. Fabbietti and P. Gasik and T. Klemenz},
  journal= {arXiv preprint arXiv:1906.06721},
  year   = {2019}
}

Comments

14 pages, 11 figures

R2 v1 2026-06-23T09:54:55.973Z