Abstract

Multi Micro-Channel-Plate Photomultiplier tubes (MCP-PMT) with High Collection Efficiency (Hi-CE) MCPs are developed and characterised. With these Hi-CE MCPs more than 90% of photoelectrons emitted from the photocathode can be detected; this is in contrast to conventional MCPs where about 50% of photoelectrons are lost at the MCP stage. The drawback of the Hi-CE MCPs is a small degradation of the transfer time spread (TTS). However for applications where no sub-ns time resolution is required the implementation of Hi-CE MCPs is extremely beneficial, as it improves the detection efficiency of the MCP-PMT almost by a factor of two.

Keywords

Cite

@article{arxiv.1711.03897,
  title  = {High collection efficiency MCPs for photon counting detectors},
  author = {D. A. Orlov and T. Ruardij and S. Duarte Pinto and R. Glazenborg and E. Kernen},
  journal= {arXiv preprint arXiv:1711.03897},
  year   = {2018}
}

Comments

8 pages, 3 figures