English

Energy filtering with X-ray lenses: optimization for photon-counting mammography

Medical Physics 2021-01-12 v1

Abstract

Chromatic properties of the multi-prism and prism-array X-ray lenses (MPL and PAL) can potentially be utilized for efficient energy filtering and dose reduction in mammography. The line-shaped foci of the lenses are optimal for coupling to photon-counting silicon strip detectors in a scanning system. A theoretical model was developed and used to investigate the benefit of two lenses compared with an absorption-filtered reference system. The dose reduction of the MPL filter was ~15% compared with the reference system at matching scan time, and the spatial resolution was higher. The dose of the PAL-filtered system was found to be ~20% lower than for the reference system at equal scan time and resolution, and only ~20% higher than for a monochromatic beam. An investigation of some practical issues remains, including the feasibility of brilliant-enough X-ray sources and manufacturing of a polymer PAL.

Keywords

Cite

@article{arxiv.2101.03494,
  title  = {Energy filtering with X-ray lenses: optimization for photon-counting mammography},
  author = {Erik Fredenberg and Bjorn Cederstrom and Mats Danielsson},
  journal= {arXiv preprint arXiv:2101.03494},
  year   = {2021}
}
R2 v1 2026-06-23T21:57:33.327Z